Help with Calibration
Check out the procedure for peace of mind to maintain your instrument’s accuracy.

The system is supplied with a factory force calibration (which can be restored at any time). However, if you wish to calibrate at a specific weight, you can perform the following procedure:
Force

1
From the tool bar menu, click on T.A. > Calibrate > Calibrate Force.

2
Select either Factory or User, as required. The installed loadcell capacity is identified.

3
Select the calibration weight of your choice.
Probe Height
Only required if you wish to: a: measure in % strain; b: record the height of the product during the test; c: use a button trigger; d: start each test from the same position; e: pre-set probe positions in the Move Probe window.